Prodrive Technologies logo white Prodrive Technologies logo
|

Contact Us

Product navigation

NG2200-XP

Premium MRI Gradient Amplifier for 3.0T ultra-high field MRI systems

Product features

  • Cost effective three-axes Gradient Amplifier cabinet for premium MRI systems
  • 2100V / ±1200Ap / 360Arms
  • Patented end stage
  • Integrated PT high precision current sensors
  • Low EMI by integrated output filter
  • Isolated power supply with PFC
  • Digital control and advanced diagnostics
  • High reliability of >30khrs.
Contact Us
NG1800 2 Resized
2100V / 1200Ap / 360Arms
Benchmarked top-of-market fidelity
Extreme reliable & lifetime over 10 years
Advanced control and diagnostics with NG Tool Suite

Configuration

labels.png

communication.png

Ngtoolsuite Icon

Specifications

Isolated power supply

  Min Typical Max Unit
Mains 380   480 VAC +/- 10%
  50   60 Hz
Mains switch included        
Power Factor Correction (PFC)     Unity power factor & low harmonics 
Programmable bus voltage 1200   2100 Vdc
Mains input power   90   kVA
Available power to end stage   65   kW per axis
    82   kW for three axis

Output performance

  Min Typical Max Unit
Output voltage (at Ipeak)    +/-1900    V
Output current   900   Apk
    360   Arms Continuous or DC
Load range 0.22   4.7 mH
   50   350
Patented end stage included         
Integrated output filter (fcorner<switching frequency)       

Output Accuracy

  Min Typical Max Unit
Settling time   error<1% after 100µs     
    error<0.4% after 150µs
    error<0.1% after 250µs
Drift offset typical    5mA/°C    
Drift gain typical    10ppm/°C    
Propagation delay  <70µs      
Linearity <100ppm      
Bandwith      >10 kHz
Total harmonic distortion <0.4% at 230Arms/200Hz      
  <0.8% at 230 Arms/1kHz      
Initial offset: neglectable (auto-zero )         
High precision current sensors included