Prodrive Technologies logo white Prodrive Technologies logo
|

Contact Us

Product navigation

NGRF1500

High performance RF Amplifier for 1.5T high-field MRI systems

Key features

  • Cost effective single channel 1.5T RF Amplifier
  • Extremely rugged design / high reliability
  • State of the art technology
  • 20kW PEP, 10% duty cycle, 8ms pulse length
  • Isolated power supply with PFC (Power Factor Correction)
  • Direct mains connection, no PDU required
  • Advanced hybrid control loop to achieve high fidelity
  • Advanced diagnostics
  • EC 60601 edition 3, CE
  • Liquid cooled
  • Ask for availability
Contact Us
NGRF1500 1.1 Left Angle
20kW PEP / 2kW AVG / 10% Duty Cycle
Direct mains connection, no PDU required
Extreme reliable & lifetime over 10 years
Advanced diagnostics & control with NG Tool Suite

Configuration

communication.png

labels.png

Ngtoolsuite Icon

Specifications

Isolated Power Supply

  Min Max Unit
Mains 380 480 VAC +/- 10%
  50 60 Hz
Mains switch included      
Power Factor Correction (PFC)      

Power Performance 

  Typical Unit
Pnom 20 kW PEP
Pav @20kW  PEP/10% Duty cycle 2 kW  
Pcw 200 W PEP
Pulse length  @ Pnom 8 ms

Gain & Frequency Range

  Typical Unit
Gain 73 dB +/-0.2dB
Gain flatness (+/-500 kHz) +/-0.2 dB
Operating frequency  63.87  +/-1MHz